The monitor is designed to measure the 132I radionuclide in the coolant of the primary circuit of a nuclear reactor, which is released into the coolant only in case of fuel cell leaks. The volumetric activity is measured using the 132I property, which emits two gamma photons of 667 and 772 keVwhen decayed. These photons are detected
simultaneously by 2 opposite GD-52 detectors, between which the measured liquid is guided in the tube.
The monitor consists of:

  • 2 × GD-52 based on 2.5" × 2.5" NaI(Tl) gamma detector (correct detector functionality can be checked by built-in test LED),
  • Sandwich shielding GD-52 (lead, tin, copper) near the pipeline with the measured liquid,
  • Stainless steel cooler GD-52, through which the coolant flows (ensuring that the maximum working temperature of the detector is not exceeded),
  • RPU-12 Data Acquisition and Processing Units with two 1024 Channel Analyzers (MCA) to evaluate data from detectors..

RELATED PRODUCTS

  • LAD-07 Liquid Activity Detector

 

 

Download PDF

GET EXPERT HELP

Specifications, standards and device design may be subject to change with regard to device development or specific customer design, without prior notice of change.
Please contact our representative to confirm the information provided on this website.

  • 60+
  • products for
    nuclear power plants
  • 4
  • different type
    of gamma irradiators
  • 3500+
  • annual distribution
    of contamination monitors
+420 516 428 611
The Czech Republic
Svitavska 588, 679 21 Cerna Hora


Certifications and authorisations